Fast Response Mass Flow Meters and Mass Flow Controllers
300-D-Series

Mass flow techniques provide more RELIABLE, REPEATABLE AND ACCURATE gas flow rate measurements compared to volumetric methods such as rotometers and turbine meters. Mass flow measurements use the density of the gas to translate mass flow units (kg or lbs) to units of familiar volumetric terms (liters/min). However the measurements are still mass units (we call those units STANDARD liter/min (slm)) and they do not change with temperature and pressure changes. We do this by referencing just one, specific temperature and one specific pressure (STP).

300-D-Series mass flow instruments have flow ranges from 0-5 sccm to 0-3,000 slm; use all metal seals, provide analog outputs, and are constructed of stainless steel. The mass flow controllers have normally closed valves. Corrosive gases, with the exception of the halogens, pressures up to 1000 psig (optional) are easily measured.

Self diagnostics of the sensor, valve and overflow are included in the digital features. Also of interest is the auto zero feature which detects when the valve is closed and automatically resets the zero. These mass flow meters and controllers can store up to 10 calibrations of different ranges and gases. When monitoring different gases in the same level of magnitude, (flow rate, molecular weight) this instrument is economically superior and more efficient when changing gases. Another feature is the built in totalizer for monitoring gas usage under user set conditions.

± (0.5% of reading + 0.2% of full scale) Accuracy for 3,000 slm

Go To Volume (Flow Rate) Conversions

Features
• +/- (0.5% of reading + 0.2% of Full Scale)
for 0 - 2,500 slm
• Proven Reliability
• Range 5 sccm to 3,000 slm(N2 Equivalent)
• NIST Traceable Calibration
• Built in Totalizer
• Automatic Zeroing

Accessories For Meters And Controllers
PowerPod Models 40, 100, & 400
Cables
Flow Service Plan

Applications
• University Research
Leak Testing
• Vapor Deposition
Pharmaceutical Processes
• R&D And Process Flows
• Semiconductor Processes
• Gas Blending
Thermal Flame Spray
• Applications where Accuracy is Critical

300-D-Series Mass Flow Meter and Controller Specification (Based on Nitrogen)
Standard Temperature and Pressure - 0° C and 760 Torr

*Compliance Data Forthcoming

Specification
Mass Flow Meter & Controller
HFM-D-300
HFC-D-302
Mass Flow Meter & Controller
HFM-D-301
HFC-D-303
Mass Flow Meter & Controller
HFM-D-305
HFC-D-307
View Product Bulletins for ordering codes Find Ordering Codes On Last Page Find Ordering Codes On Last Page Find Ordering Codes On Last Page
Manuals HFM-D-300 /
HFC-D-302
HFM-D-301 /
HFC-D-303
HFM-D-305 /
HFC-D-307
Drawings HFM-D-300 /
HFC-D-302
HFM-D-301 /
HFC-D-303
HFM-D-305 /
HFC-D-307
Full Scale Flow Range 0-5 sccm to
0-25,000 sccm
0-25 slm to
0-1000 slm
0-1000 slm to
0-3,000 slm
Accuracy +/- (0.5% of reading + 0.2% of Full Scale) +/- (0.5% of reading + 0.2% of Full Scale) +/- (0.5% of reading + 0.2% of Full Scale)
+/-2% of full scale for 2,500 - 3,000 slm
Repeatability +/-0.1% of Full Scale +/-0.1% of Full Scale +/-0.1% of Full Scale
Standard Pressure Rating 500 psig 500 psig 500 psig
Pressure Coefficient +0.01%/psi of reading (0-50 psig N2) typical
+0.01% of reading/psi (0-1000 psig N2) typical +0.015%/psi(0-500 psig N2)
High-Pressure Option 1000 psig (proof tested to 1500 psig) 1000 psig Not Available
Temperature Coefficient of Meter Zero <0.2%/o C of Full Scale
Span <0.05%/ o C of Reading
Zero <0.2%/o C of Full Scale
Span <0.05%/ o C of Reading
Zero <0.2%/o C of Full Scale
Span <0.05%/ o C of Reading
Temperature Coefficient of Controller Zero insignificant when auto zero is active
Span <0.05%/ o C of Reading
Zero insignificant when auto zero is active
Span <0.05%/ o C of Reading
Zero insignificant when auto zero is active
Span <0.05%/ o C of Reading
Leak Integrity < 1 x 10-9 sccs He 1 x 10-9 sccs He 1 x 10-9 sccs He
Power - Meter +/-15VDC at 55 mA +/-15VDC at 55 mA +/-15VDC at 55 mA
Power - Controller +/-15 VDC at 150 mA +/-15 VDC at 150 mA +/-15 VDC at 150 mA
Flow Signal 0-5.00 VDC, 0-10 VDC
0-20 mAmp, 4-20 mAmp
0-5.00 VDC, 0-10 VDC
0-20 mAmp, 4-20 mAmp
0-5.00 VDC, 0-10 VDC
0-20 mAmp, 4-20 mAmp
Attitude Sensitvity of Zero < 0.7% of Full Scale (N2@19.7 psia) < 0.7% of Full Scale < 0.7% of Full Scale
Attitude Sensitvity of Span < 0.05% of Full Scale (N2@19.7 psia) < 0.05% of reading < 0.05% of reading
Settling Time of Meter </= 0.5 sec (0% to 100% FS) </=0.5 sec (0% to 100% FS) </=0.5 sec (0% to 100% FS)
Settling Time of Controller </= 0.75 sec (0% to 100% FS) </= 2.0 sec (0% to 100% FS) </= 2.0 sec (0% to 100% FS)
Wetted Material - Meter 302 SS, 316 LSS, Ni 200 Viton® , 302 SS, 316 LSS, Ni 200 302 SS, 316 LSS, NI 200
Wetted Material - Controller 302 SS, 316 LSS, Ni 200, Kalrez ® , (Valve Seat) Viton® , 302 SS, 316 LSS, Ni 200, Teflon ® ,(Valve Seat) Kalrez® 302 SS, 316 LSS, Ni 200, Kalrez ® (Valve Seat)
Connector 15-pin subminiature D 15-pin subminiature D 15-pin subminiature D
Digital Communication RS 232 (STD)
RS 485
RS 232 (STD)
RS 485
RS 232 (STD)
RS 485
Pin out diagrams H - Pin out H - Pin out H - Pin out
Fittings 1/4" Swagelok® (STD) Options
1/8" Swagelok®
1/4" VCR® pair
1/4" VCO® pair
6mm Swagelok® pair
No Fittings
1/2" Swagelok® (STD) Options
1/2" VCR® pair
1/2" VCO® pair
12mm Swagelok® pair
No Fittings
1" Swagelok® (STD) Options
1" VCR® pair
1" VCO® pair
25mm Swagelok® pair
No Fittings
Weight (approx) HFM 300 - 1.95 lbs
HFC 302 - 2.45 lbs
HFM 301 - 3.5 lbs
HFC 303 - 5.3 lbs
HFM 305 - 8.2 lbs
HFC 307 - 15.3 lbs

 

 



Home | ISO 9001 | Terms of Use | Careers | Site Map | Contact

Teledyne Hastings Instruments 804 Newcombe Ave, Hampton, VA 23669
757.723.6531 | http://www.teledyne-hi.com | [F] 757.723.3925