Mass flow techniques provide more RELIABLE, REPEATABLE AND ACCURATE gas flow rate measurements compared to volumetric methods such as rotometers and turbine meters. Mass flow measurements use the density of the gas to translate mass flow units (kg or lbs) to units of familiar volumetric terms (liters/min). However the measurements are still mass units (we call those units STANDARD liter/min (slm)) and they do not change with temperature and pressure changes. We do this by referencing just one, specific temperature and one specific pressure (STP).
300-D-Series mass flow instruments have flow ranges from 0-5 sccm to 0-3,000 slm; use all metal seals, provide analog outputs, and are constructed of stainless steel. The mass flow controllers have normally closed valves. Corrosive gases, with the exception of the halogens, pressures up to 1000 psig (optional) are easily measured.
Self diagnostics of the sensor, valve and overflow are included in the digital features. Also of interest is the auto zero feature which detects when the valve is closed and automatically resets the zero. These mass flow meters and controllers can store up to 10 calibrations of different ranges and gases. When monitoring different gases in the same level of magnitude, (flow rate, molecular weight) this instrument is economically superior and more efficient when changing gases. Another feature is the built in totalizer for monitoring gas usage under user set conditions.
Specification |
Mass Flow Meter & Controller
HFM-D-300
HFC-D-302 |
Mass Flow Meter & Controller
HFM-D-301
HFC-D-303 |
Mass Flow Meter & Controller
HFM-D-305
HFC-D-307
|
| View Product Bulletins for ordering codes |
Find Ordering Codes On Last Page |
Find Ordering Codes On Last Page |
Find Ordering Codes On Last Page |
| Manuals |
HFM-D-300 /
HFC-D-302 |
HFM-D-301 /
HFC-D-303 |
HFM-D-305 /
HFC-D-307 |
| Drawings |
HFM-D-300 /
HFC-D-302 |
HFM-D-301 /
HFC-D-303 |
HFM-D-305 /
HFC-D-307 |
| Full Scale Flow Range |
0-5 sccm to
0-25,000 sccm |
0-25 slm to
0-1000 slm |
0-1000 slm to
0-3,000 slm |
| Accuracy |
+/- (0.5% of reading + 0.2% of Full Scale) |
+/- (0.5% of reading + 0.2% of Full Scale) |
+/- (0.5% of reading + 0.2% of Full Scale)
+/-2% of full scale for 2,500 - 3,000 slm |
| Repeatability |
+/-0.1% of Full Scale |
+/-0.1% of Full Scale |
+/-0.1% of Full Scale |
| Standard Pressure Rating |
500 psig |
500 psig |
500 psig |
| Pressure Coefficient |
+0.01%/psi of reading (0-50 psig N2) typical |
+0.01% of reading/psi (0-1000 psig N2) typical |
+0.015%/psi(0-500 psig N2) |
| High-Pressure Option |
1000 psig (proof tested to 1500 psig) |
1000 psig |
Not Available |
| Temperature Coefficient of Meter |
Zero <0.2%/o C of Full Scale
Span <0.05%/ o C of Reading |
Zero <0.2%/o C of Full Scale
Span <0.05%/ o C of Reading |
Zero <0.2%/o C of Full Scale
Span <0.05%/ o C of Reading |
| Temperature Coefficient of Controller |
Zero insignificant when auto zero is active
Span <0.05%/ o C of Reading
|
Zero insignificant when auto zero is active
Span <0.05%/ o C of Reading |
Zero insignificant when auto zero is active
Span <0.05%/ o C of Reading |
| Leak Integrity |
< 1 x 10-9 sccs He |
1 x 10-9 sccs He |
1 x 10-9 sccs He |
| Power - Meter |
+/-15VDC at 55 mA |
+/-15VDC at 55 mA |
+/-15VDC at 55 mA |
| Power - Controller |
+/-15 VDC at 150 mA |
+/-15 VDC at 150 mA |
+/-15 VDC at 150 mA |
| Flow Signal |
0-5.00 VDC, 0-10 VDC
0-20 mAmp, 4-20 mAmp |
0-5.00 VDC, 0-10 VDC
0-20 mAmp, 4-20 mAmp |
0-5.00 VDC, 0-10 VDC
0-20 mAmp, 4-20 mAmp |
| Attitude Sensitvity of Zero |
< 0.7% of Full Scale (N2@19.7 psia) |
< 0.7% of Full Scale |
< 0.7% of Full Scale |
| Attitude Sensitvity of Span |
< 0.05% of Full Scale (N2@19.7 psia) |
< 0.05% of reading |
< 0.05% of reading |
| Settling Time of Meter |
</= 0.5 sec (0% to 100% FS) |
</=0.5 sec (0% to 100% FS) |
</=0.5 sec (0% to 100% FS) |
| Settling Time of Controller |
</= 0.75 sec (0% to 100% FS) |
</= 2.0 sec (0% to 100% FS) |
</= 2.0 sec (0% to 100% FS) |
| Wetted Material - Meter |
302 SS, 316 LSS, Ni 200 |
Viton® , 302 SS, 316 LSS, Ni 200 |
302 SS, 316 LSS, NI 200 |
| Wetted Material - Controller |
302 SS, 316 LSS, Ni 200, Kalrez ® , (Valve Seat) |
Viton® , 302 SS, 316 LSS, Ni 200, Teflon ® ,(Valve Seat) Kalrez® |
302 SS, 316 LSS, Ni 200, Kalrez ® (Valve Seat) |
| Connector |
15-pin subminiature D |
15-pin subminiature D |
15-pin subminiature D |
| Digital Communication |
RS 232 (STD)
RS 485 |
RS 232 (STD)
RS 485 |
RS 232 (STD)
RS 485 |
| Pin out diagrams |
H - Pin out |
H - Pin out |
H - Pin out |
| Fittings |
1/4" Swagelok® (STD) Options
1/8" Swagelok®
1/4" VCR® pair
1/4" VCO® pair
6mm Swagelok® pair
No Fittings
|
1/2" Swagelok® (STD) Options
1/2" VCR® pair
1/2" VCO® pair
12mm Swagelok® pair
No Fittings |
1" Swagelok® (STD) Options
1" VCR® pair
1" VCO® pair
25mm Swagelok® pair
No Fittings |
| Weight (approx) |
HFM 300 - 1.95 lbs
HFC 302 - 2.45 lbs |
HFM 301 - 3.5 lbs
HFC 303 - 5.3 lbs |
HFM 305 - 8.2 lbs
HFC 307 - 15.3 lbs |
*Accuracy is for Nitrogen. See manuals for Gas Conversion Factors